Technique
1. For a convenient manipulation of the CerclageFix, use the Screw
Holding Sleeve for LCP 4.5/5.0 (314.281) to capture the CerclageFix.
2. Insert the CerclageFix in the selected plate hole using the Screw
Holding Sleeve.
3. The CerclageFix does not have to be locked in the plate. Align the hole
of the CerclageFix with the desired position of the wire.
4. Place the plate on the bone.
5. Insert the screws through the plate using standard AO/ASIF tech-
niques.
6. Note that the CerclageFix for LCP 4.5/5.0 can also be inserted onto the
plate after fixing the plate to the bone.
7. Pass the cerclage wire through the CerclageFix hole and around the
bone.
8. Twist the wire to secure it.
Order information
481.002 CerclageFix for LCP 4.5/5.0, pure titanium
281.002 CerclageFix for LCP 4.5/5.0, stainless steel
Also available
107.920 Wire Instrument Set in Sterilizing Tray
105.920 Wire Instrument Set in Graphic Case
188.215 Cable System in Vario Case
291.270–291.280 Cerclage Wires without eye, stainless steel
291.010–291.120 Cerclage Wires with eye, stainless steel